Chapter 5 149
Analog Measurements - Screens and Control Fields
Scope To Scope To
Scope To
This field selects the signal source for the oscilloscope. This allows you
to bypass certain sections of the AF analyzer’s circuitry when viewing
and measuring a signal. It also allows you to select measurement paths
that include additional gain stages, improving the oscilloscope’s
resolution when measuring low-level signals.
Input looks at the unfiltered signal directly from the input.
Filters looks at the signal after filtering through Filter 1 and
Filter 2.
De-Emp looks at the signal after filtering through Filter 1 and
Filter 2, and after 750 µs de-emphasis (if the De-emphasis field is
set to 750 us.)
Notch looks at the signal after filtering through Filter 1 and Filter
2, after de-emphasis (if used), and after the notch filter.
Screens on which this field is displayed
AF ANALYZER
Main Menu