HP (Hewlett-Packard) E6380-90019 Cell Phone Accessories User Manual


 
162 Chapter 6
Configuration - Screens and Control Fields
Abort Print Antenna In
Control Fields on the Configuration Screens
Abort Print
Select this field to interrupt the printing job in progress.
Abort Print is not directly programmable over the HP-IB.
Screen(s) Where Field is Present
PRINTER CONFIGURE
Antenna In
This field is used to indicate losses or gains between the ANT IN port
and the device-under-test.
Enter a positive value to indicate a gain (such as an amplifier). The
spectrum analyzer’s marker level (Lvl) measurement is automatically
reduced by that amount. The spectrum analyzer’s Ref Level is
automatically decreased by the same amount, so the trace position does
not appear to change.
Enter a negative value to indicate a loss (such as cable loss). The
spectrum analyzer marker’s level (Lvl) measurement is automatically
increased by that amount. The spectrum analyzer’s Ref Level is
automatically increased by the same amount, so the trace position does
not appear to change.
This field is only used when the RF Level Offset field is set to On
.
See "RF Level Offset" on page 178.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
Main Menu