HP (Hewlett-Packard) E6380-90019 Cell Phone Accessories User Manual


 
Chapter 6 173
Configuration - Screens and Control Fields
Range Hold Range Hold
Range Hold
These fields enable/disable several autoranging and autotuning
routines.
Auto All enables these routines, providing automatic adjustment
when making AF or RF measurements.
Hold All disables these routines, requiring you to manually set the
affected settings.
The following fields are affected by the Range Hold field:
Input Atten in the RF ANALYZER and SPEC ANL screens.
Gain Cntl in the AF ANALYZER screen. This field controls three
AF gain setting fields:
Input Gain
De-Emp Gain
Notch Gain
Operating Considerations
The Hold All setting is primarily used when the instrument is
operated by remote control, such as in an automated test system.
Unless you have very specific reasons for disabling the automatic
functions, you should set this field to Auto All when operating the
instrument manually.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
Main Menu