7.3 CALIBRATION
Next is the Calibration section. “WEl” will be displayed.
When WEI is displayed, press the [Unit] key to scroll through the
options for selecting the calibration mass.
To select the desired mass press the [Mode] key.
The raw ADC count will be displayed along with “CAI”.
To set the zero calibration, press the [Mode] key again.
The scale will ask you to load the calibration mass by displaying the
new ADC and “LOD
Place the calibration mass (same as the chosen WEI capacity) and
when the stable indicator is on, press [Mode] again.
The display will show the value of the calibration mass.
Remove the weight.
The scale will return to the normal weighing.