A SERVICE OF

logo

98
air bubbles remain trapped in the sleeve.
Record the temperature of the buffer solution from the thermom-
eter (e.g. 18°C).
Switch the instrument on by pressing ON/OFF.
Press COND/TEMP (HI 8633) or
TDS/TEMP (HI 8734) to display
the temperature.
Adjust the TEMPERATURE knob to display 18°C.
Press COND/TEMP (or TDS/TEMP)
again to display conductivity mea-
surement.
Select 19.99 mS/cm (HI 8633)
or 1999 mg/L (HI 8734) range
by pressing the appropriate
range key.
Remove the battery compartment cover and adjust the calibration
trimmer located inside with the calibration screwdriver until the
display shows for HI 8633 the conductivity reading at 25°C (see
the conductivity vs. temperature table), e.g. @ 25°C, 12880 µS/
cm = 12.88 mS/cm, or for HI 8734 the TDS reading at 25°C
(see the TDS vs. temperature chart), e.g. @25°C, 1382 mg/L.
ON
OFF
TEMPERATURE
COND.
TEMP.
TDS
TEMP.
COND.
TEMP.
TDS
TEMP.
19.99
mS
1999
mg/l
Select the appropriate measurement range.
Note: If the display shows only a "1" on the
far left hand side, the meter is out of
range. Select the next (higher) range.
• Wait for a couple of minutes for the temperature sensor to reach
thermal equilibrium with the sample before taking measure-
ments.
After the measurement has been completed, the instrument
should be switched off and the probe should be cleaned and
dried (see "Probe Maintenance" on page 15).
Accessories needed:
Use any calibration solution within the meter's range. The
solution should ideally be close to the samples being mea-
sured. Use for example HI 7030 or HI 8030, 12880 µS/cm
(=12.88 mS/cm) conductivity solution, for HI 8633 and HI
8733, and HI 7032, 1382 mg/L (=2764 µS/cm) TDS
solution, for HI 8734.
ChecktempC or another accurate thermometer with 0.1°C
resolution (not necessary for HI 8733).
a small screwdriver.
PROCEDURE FOR HI 8633 AND HI 8734
Pour sufficient quantity of a conductivity
(HI 8633) or TDS (HI 8734) calibration
solution (e.g. HI 7030 or HI 7032) into
a beaker to cover the holes on the probe.
If possible, use plastic beakers to mini-
mize any EMC interference.
Immerse the conductivity probe, making
sure that holes are completely submerged,
and the ChecktempC in the solution.
Wait for a couple of minutes for thermal
equilibrium to be reached.
Tap the probe on the bottom, then
shake it while rotating to make sure no
CALIBRATION
HI 7030
°C