Model 5328A
General Information
Table 1-3. 5328A Counter Specifications
GENERAL
Power Requirements:
115 or 230 volts 60 or
400 Hz ac.
Display:
Nine-digit LED.
Sample Rate:
Variable from less than 2 milli-
seconds to HOLD.
Arming:
Rear panel ARM (ON-OFF) switch.
Refer to operation for details.
Blanking:
Unwanted zeros to left of most-
significant-digit are suppressed.
Hold:
HOLDS count between samples.
Trigger Light:
Indicates input is above trigger
level.
PROGRAMMABLE OPERATION
(CHANNELS A AND B)
Includes independent selection of coupling,
trigger slope, trigger level, and attenuator for
each channel. Separate/Common A switch is
programmable. Also, an invert feature switches
Channels A and B; useful in all functions except
Ratio B/A.
Trigger level is programmable in 10 mV steps
in X1; 100 mV in X10; IV in X100.
Trigger level accuracy under remote control:
X1:
±35 mV
X10:
±350 mV +2% of trigger level
X100:
±3.5V +2% of trigger level
Input Characteristics
Sensitivity:
15 mV rms, 0—35 MHz (decoupled
20 Hz—35 MHz (ac coupled)
50 mV rms, 35 MHz—100 MHz
Minimum pulse width 5 ns, 140 mV p-p.
Coupling:
ac or dc switch selectable.
Impedance:
1
MO
II <70 pF.
Trigger Level:
Variable over ±2.5 volts times
attenuator setting with 0 volt preset position.
Trigger Slope:
Independent selection of + or
- slope.
Attenuators:
X1, X10, X100.
Dynamic Range:
25 mV to 1V rms times
attenuator setting, 0—35 MHz; 50 mV rms
times attenuator setting, 35 MHz to 100 MHz
Maximum Input:
dc coupled, X1:
250V rms, dc — SO kHz
1.25 x 10
7
V rms/freq., 50 kHz—2.5 MHz
5V rms, 2.5 MHz—100 MHz.
dc coupled, X10 and X100:
250V rms, dc 5
MHz 1.25 x 109V rms/freq., 5—100 MHz
ac coupled:
200V (peak ac + dc),
0-20 Hz; same as dc coupled above 20 Hz.
Channel Input:
Separate or Common A.
Marker Outputs:
A and B channel Schmidt
trigger outputs available on front panel;
0 to 300 mV levels into
50Q;
<20 ns delay.
Frequency Measurements
Frequency A
Range:
0—100 MHz direct count.
Resolution:
1 MHz to 0.1 Hz in decade steps.
Accuracy:
±1 count ± time base error.
Display: Hz, kHz, MHz.
Period Measurements
Period A
Range:
0—10 MHz
Resolution:
10 ns to 0.1s in decade steps.
Accuracy:
±1 count ± time base error
±trigger error*
Display:
ns,
ps,
ms.
Period Average A
Range:
0—10 MHz
Resolution:
100 ns—.01 ps in decade steps.
Accuracy:
±1 count displayed ± time base
error
trigger error*
±
no. of periods averaged
Display:
Ps,
ns
Time Interval Measurements
Time Interval A to B
Range:
100 ns to 10
8
seconds
Resolution:
100 ns to 1-second in decade
steps.
Accuracy:
±1 count ± time base error
± trigger error*
Display:
ps,
ms, s.
Time Interval Average A to B
Range:
0.1 ns to 10 seconds
Resolution:
Accuracy:
Minimum Dead Time:
150 ns from one
STOP to next START
Maximum Repetition Rate:
10 MHz.
Display:
Ps,
ns.
Ratio Measurement
Ratio B/A, or C/A
Range:
A: 0—10 MHz
Range:
B: 0—100 MHz
Range:
C: 30—500 MHz
Resolution:
1 part in
~
x N
Accuracy: ±1 count of B or C ± trigger
error* of A times frequency of B or C (N>1)
For N=1, add 12 ns times frequency of
B or C.
* Trigger error is <0.3% of one period for sine waves
of 40 dB S/N or better and amplitude equal to
sensitivity of counter. For any wave shape, trigger
error is than then:
1-3