THS710A, THS720A, THS730A & THS720P Service Manual
4–1
Performance Verification
This chapter contains performance verification procedures for the specifications
marked with the n symbol. The following equipment, or a suitable equivalent,
is required to complete these procedures.
Description
Minimum
Requirements
Examples
DC Voltage Source 60 mV to 800 V, ±0.1% accu-
racy
av
t
k
niv
al
Calibration System with Oscil-
AC Voltage Source 300 mV to 640 V, ±0.5%
accuracy at 500 Hz
lo
co
ali
ation Mo
ul
(Option 250)
Resistance Standard 360 W to 36 MW, ±0.1%
accuracy
luk
Multi
o
uct
Calibrator with Oscilloscope
ali
ation
tion (
tion
Leveled Sine Wave Generator 50 kHz and 100 MHz, ±3%
amplitude accuracy
ali
ation
tion (
tion
5500A-SC)
Time Mark Generator 10 ms period, ±50 ppm accu-
racy
Banana to Banana Cable (two
required)
Shielded banana jacks on
each end
Tektronix Deluxe Meter Lead
Set (ATLDX1)
50 W BNC Cable BNC male to BNC male,
≈ 36 in (1 m) long
Tektronix part number
012-0482-00
50 W Feedthrough Termina-
tion
BNC male and female con-
nectors
Tektronix part number
011-0049-01
Dual Banana to BNC Adapter Banana plugs to BNC female Tektronix part number
103-0090-00