A SERVICE OF

logo

Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 101
Triggering Events and Data Parameters/Functions for these default channels are:
LIST OF CHANNELS LIST OF PARAMETERS
NAME:
CONC
EVENT:
ATIMER
REPORT PERIOD:
000:01:00
NO. OF RECORDS:
800
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
ON
PARAMETER:
CONC1
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
NAME:
PNUMTC
EVENT:
ATIMER
REPORT PERIOD:
001:00:00
NO. OF RECORDS:
360
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
PARAMETER:
SMPLFLW
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
SMPLPRS
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
NAME:
CALDAT
EVENT:
SLPCHG
REPORT PERIOD: N/A
NO. OF RECORDS:
200
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
PARAMETER:
OFSET1
MODE:
INST
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
SLOPE1
MODE:
INST
PRECISION:
3
STORE NUM SAMPLES
OFF
PARAMETER:
ZSCNC1
MODE:
INST
PRECISION:
1
STORE NUM SAMPLES
OFF
NAME:
STBZRO
EVENT:
EXITZR
REPORT PERIOD: N/A
NO. OF RECORDS:
200
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
PARAMETER:
DETMES
MODE:
INST
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
STABIL
MODE:
INST
PRECISION:
2
STORE NUM SAMPLES
OFF
PARAMETER:
RATIO
MODE:
INST
PRECISION:
3
STORE NUM SAMPLES
OFF
NAME:
STBSPN
EVENT:
EXITSP
REPORT PERIOD: N/A
NO. OF RECORDS:
200
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
PARAMETER:
DETMES
MODE:
INST
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
RATIO
MODE:
INST
PRECISION:
3
STORE NUM SAMPLES
OFF
NAME:
TEMP
EVENT:
ATIMER
REPORT PERIOD:
000:06:00
NO. OF RECORDS:
400
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
PARAMETER:
BOXTMP
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
BNTEMP
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
PHTDRV
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
Figure 6-7: Default iDAS Channels Setup