51
• Attach one end of stinger to shaker and one end to force transducer
• Connect amplifier output to shaker
• Connect amplifier input to AGILENT 35670A source
Connect the transducers
• Connect force transducer to Channel 1
• Connect accelerometer to Channel 2
Preset the AGILENT 35670A
• Turn on the AGILENT 35670A, or
• Press the green [Preset] hardkey
• Press the yellow [Pause/Cont] hardkey if desired.
Enter Channel 1 input parameters
• Press the [Input] hardkey
• Press [CH1 AUTO UP ONLY] softkey (F5).
• Press [FRONT END CH1 SETUP] softkey (F7).
• Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.
• Press the [Input] hardkey
• Press [XDCR UNIT CH1 SETUP] softkey (F8)
• Press the [XDCR SENSITVTY] softkey (F6).
• Enter 8, then press the [mV/EU] softkey (F2).
• Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.
• Press the [XDCR UNIT LABEL] softkey (F7).
• Press the [MORE CHOICES] softkey (F9).
• Select the [lb] softkey (F6).
Enter Channel 2 input parameters
• Press the [Input] hardkey
• Press [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.
• Press [CH2 AUTO UP ONLY] softkey (F5).
• Press [FRONT END CH2 SETUP] softkey (F7).
• Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.
• Press the [Input] hardkey
• Press [XDCR UNIT CH2 SETUP] softkey (F8)
• Press the [XDCR SENSITVTY] softkey (F6).
• Enter 10.2, then press the [mV/EU] softkey (F2).
• Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.
• Press the [XDCR UNIT LABEL] softkey (F7).
• Select the [g] softkey (F2).