A SERVICE OF

logo

Appendix A
Stimulation current display in the mA range
Patient current Measurement display Controller display
0.00 mA 0.00 mA 0.00 mA
0.05 mA 0.05 mA ± 0.02 0.05 mA ± 0.03
0.10 mA 0.10 mA ± 0.02 0.10 mA ± 0.03
0.20 mA 0.20 mA ± 0.02 0.20 mA ± 0.03
0.30 mA 0.30 mA ± 0.02 0.30 mA ± 0.03
0.40 mA 0.40 mA ± 0.02 0.40 mA ± 0.03
0.60 mA 0.60 mA ± 0.02 0.60 mA ± 0.03
0.80 mA 0.80 mA ± 0.02 0.80 mA ± 0.03
1.00 mA 1.00 mA ± 0.02 1.00 mA ± 0.03
Patient current Measurement display Controller display
0.00 mA 0.00 mA 0.00 mA
0.50 mA 0.50 mA ± 0.1 0.50 mA ± 0.15
1.00 mA 1.00 mA ± 0.1 1.00 mA ± 0.15
1.50 mA 1.50 mA ± 0.1 1.50 mA ± 0.15
2.00 mA 2.00 mA ± 0.1 2.00 mA ± 0.15
3.00 mA 3.00 mA ± 0.1 3.00 mA ± 0.15
4.00 mA 4.00 mA ± 0.1 4.00 mA ± 0.15
5.00 mA 5.00 mA ± 0.1 5.00 mA ± 0.15
24
Appendix B
Conversion table: Amplitude (mA) and pulse width (ms) to charge (nano-Coulomb, nC)
Amplitude Pulse width Charge Pulse width Charge Pulse width Charge
mA ms nC ms nC ms nC
0.1 0.1 10 0.3 30 1.0 100
0.2 0.1 20 0.3 60 1.0 200
0.3 0.1 30 0.3 90 1.0 300
0.4 0.1 40 0.3 120 1.0 400
0.5 0.1 50 0.3 150 1.0 500
0.6 0.1 60 0.3 180 1.0 600
0.7 0.1 70 0.3 210 1.0 700
0.8 0.1 80 0.3 240 1.0 800
0.9 0.1 90 0.3 270 1.0 900
1.0 0.1 100 0.3 300 1.0 1000
1.2 0.1 120 0.3 360 1.0 1200
1.6 0.1 160 0.3 480 1.0 1600
1.8 0.1 180 0.3 540 1.0 1800
2.0 0.1 200 0.3 600 1.0 2000
2.5 0.1 250 0.3 750 1.0 2500
3.0 0.1 300 0.3 900 1.0 3000
3.5 0.1 350 0.3 1050 1.0 3500
4.0 0.1 400 0.3 1200 1.0 4000
4.5 0.1 450 0.3 1350 1.0 4500
5.0 0.1 500 0.3 1500 1.0 5000