![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/d/8b/d8b64d2c-3d06-46af-90f6-a15c61615d3b/d8b64d2c-3d06-46af-90f6-a15c61615d3b-bg24.png)
No. 99MBA038A
4-10
No. Operation Screen display and remarks
4
Selecting the command for setting the tolerance
judgment function to ON or OFF
Press the numeric key corresponding to
.
–
5
Setting the tolerance judgment function to ON
Confirm that “Tolerance Judgment” is set to “ON”,
and then press the [F5] function key corresponding
to
.
<Remarks>
The status display changes as follows.
(OFF) (ON)
6
Measurement function selection
Press
.
–
7
Measuring circle (A)
Measure circle (A) at three points.
3
(A)
2
1
<Remarks>
Follow the information displayed on the screen
to perform the measurement.
8
Inputting tolerance data
Move the cursor to the input columns and enter the
nominal values and tolerance values using the
numeric keys. Note that items with no input data are
not judged.
Output Item Nominal U. Tol. L. Tol.
Coord:
X
6.000 0.010 -0.010
Coord: Y 6.000 0.010 -0.010
Dia: D 4.000 0.010 -0.010
9
Completion of inputting the tolerance data
Press the [F5] function key corresponding to
.
–
10
Results display
The measurement results for circle (A) are
displayed on the screen.
D
(A)
Y
X
N0001
X = 6.001
Y = 6.005
D1 = 4.018 +NG
<Remarks>
If the deviation exceeds the tolerance zone,
“+NG” or “-NG” is displayed as GO/NG
judgment result.