![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/3/e6/3e68ced7-e780-4a08-9488-535798d1c4ed/3e68ced7-e780-4a08-9488-535798d1c4ed-bg7c.png)
4 - 24 COMMAND REFERENCE
•
Cancels or aborts any instrument-dependent action.
•
Cancels the effect of the
*
OPC command and the
*
OPC? query.
•
Sets the TRIGger subsystem into its IDLE state.
The
*
RST command does not affect the following:
•
State of the IEEE 488.1 interface.
•
GPIB (IEEE 488.1) address of the instrument.
•
Contents of the Output Queue.
•
Contents of the Error/Event Queue.
•
Service Request Enable setting in the SRE register.
•
Transition filters in the status subsystem.
•
Event registers in the status subsystem.
•
Event enable registers in the status subsystem.
•
Calibration data that affects the device specifications.
•
Version number set by the SYSTem:VERSion command.
•
Contents of the internal memory registers (
*
SAV/
*
RCL).
Example:
Send
→ *RST
Front panel compliance:
All settings not mentioned in the description are set according to the front panel
fixed setup, which can be recalled by pressing the keys STATUS and TEXT OFF
at the same time.
FUNCTION: DEFAULT SETTING(S):
Acquisition length 512 (samples of 16 bits)
Autoranging attenuators OFF