![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/3/e6/3e68ced7-e780-4a08-9488-535798d1c4ed/3e68ced7-e780-4a08-9488-535798d1c4ed-bgb5.png)
COMMAND REFERENCE 4 - 81
SENSe:SWEep:PDETection[:STATe]
Syntax:
SENSe:SWEep:PDETection[:STATe] <Boolean>
Query form:
SENSe:SWEep:PDETection[:STATe]?
Response:
0 | 1
0 Peak detection switched off.
1 Peak detection switched on.
Description:
Switches peak detection on or off. If peak detection is switched on, the MTB range
is limited to sequential sampling from 250 nanoseconds through 200 seconds per
division (for MTB ranges, refer to the SENSe:SWEep:TIME command). For
limitations on the peak detection speed (width of a glitch), refer to the function
PEAK DETECTION of chapter 5 in the Operating Guide. In the analog mode, the
error message -221 "Settings conflict;Digital mode required" is generated.
After a
*
RST command, peak detection is switched off.
Example:
Send
→ CONFigure:PTPeak Configures for Peak-To-Peak.
Send
→ INITIate:CONTinuous ON Sets Auto run mode.
Send
→ DISPlay:MENU MEASure Displays MEASURE menu.
Send
→ SYSTem:KEY 2 Sets MEAS1 on.
Send
→ SENSe:SWEep:PDETection on Sets peak detection on.
Send
→ DISPlay:WINDow:TEXT1:DATA? Queries MEAS1 data.
Read
← <pkpk,9438E-03,V>
Front panel compliance:
The SENSe:SWEep:PDETection command is the remote equivalent of the front
panel ACQUIRE - PEAK DET on/off softkey menu.