4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual
20 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99
A
D
C
B
View of Junction Box Head
with Cap Removed
WHITE
BLACK
RED
GREEN
Cell
Temp.
Comp.
B
D
C
A
Internal Cell Assembly
Configuration
WKSHG R
Coax Cable Shield
Conductivity/Resistivity
Analyzer
A
n a l y ze r I np u t C on n e c ti o n s
B
D
A
C
1000 ft. max.
Note 3
Note 2
Four Point Terminal Plate
with #6-32 Screw Terminals
Ce l l A s s e mb l y Co n n e ct i o n s
4905
I II III IV
V
VI
- - 333 - X1
4973
4974
IIIIII
IV
- 333 - X1 -
V
4908
I II III IV V VI
- - 333 - X1
4909
VII
Five Point Terminal Board.
Each Terminal Will Accept
#16 Gage Max. Wire
Note 1
GND
NOTES:
1. For pure water samples in non-conductive (plastic, glass, etc.) piping, ground the black cell electrode lead near the cell. Alternatively,
connect to the 7082 ground screw as shown dotted. Do not ground 10, 25, or 50 constant cells.
2. 7
0 8 2- 1 6, 17 , 1 8 , 1 9 (only)
Use 22 gage minimum coaxial cable type RG59/U connecting shield to terminal "SH" only.
3. 7
0 8 2 -1 6 , 1 7 , 1 8 , 1 9
For cable runs of up to 500 ft., use: 18 gage minimum, three conductor cable.
For cbale runs of 500 - 1000 ft., use: 16 gage minimum, three conductor cable.
7
08 2 -1 3 , 14 , 1 5 [coax and shield (SH) not used]
For cable runs of up to 500 ft., use: 18 gage minimum, four conductor cable.
For cable runs of 500 - 1000 ft., use: 16 gage minimum, four conductor cable.
4. Cell to analyzer cables are considered low level. Run seperate from high level wiring.
a
/
n 23345
Figure 5-1 Installation Diagram-Cat. 4909 Cells with Junction Box Head
Connected to 7082 Conductivity/Resistivity Analyzer