4-112
Screens
RF Generator / RF Analyzer (RF Analyzer)
RF Generator / RF Analyzer (RF Analyzer)
1. Accuracy This field selects the input accuracy of the RF Amplitude setting to be assumed by
the instrument for setting the Open/Auto DAC Value when in Auto AGC mode.
Choices +-3 dB
+-1 dB
2. AGC Mode This field selects the AGC Mode.
Choices Closed is closed-loop AGC operation. It is used for stable,
repeating RF signals and provides greater accuracy measurements.
Open is used for isolated pulses, for example RACHs.
Auto is used to automatically set the Open/Auto DAC Value based
on the entered amplitude and accuracy.