A SERVICE OF

logo

6
6 - 1
No. 99MBC032A
This chapter describes the setup method and functions provided with the
RS-232C interface and Digimatic output interface of this unit.
INTERFACE UNIT
6.1 RS-232C Interface
The standard RS-232C interface of this Display Unit allows the LSM to communicate with
external devices via RS-232C (EIA standard) serial signals.
Prior to using this interface , set up the baud rate, data bits, and parity check, etc. according
to Section 4.1, “Basic Setup”. The setting contents must be compatible to that on the external
device to be connected.
6.1.1 Specifications
Applicable plug connector: D-sub 9 pin (female) (Manufacturer: AMP, HD-20/747951-1)
or equivalent.
6
9
1
5
The pin numbering for this system is shown at the left
Communication specifications
Device definition
Communication method
Synchronizing method
Baud rate
Data configuration
Specify the LSM as a terminal (DTE)
Full-duplex
Start/stop method (asynchronous)
1200, 2400, 4800, 9600, 19200 bps
Transmission code
Data bits
Start bit
Stop bit
Parity check
Delimiter
ASCII
7 or 8 bits
1 bit
1 bit
None, odd, or even
CR+LF, CR, LF