4.1.2.1 Selecting and setting the function in the B0 mode
a. Setting the resolution (Guidance: )
Set the resolution of the Measuring Unit. The resolutions that can be set for the Measuring
Units are given in “Table 4.5.2.1A” and “Table 4.5.2.1B”.
PROG
Step 1: Each time the key is pressed the displayed
setup option (number) changes in the following
order: → → ... → → → . If the
desired option is flashing, press the
ENT
key. If
the resolution setting has been made, the operation
automatically proceeds to the setting for the
number of blanked out digits.
The initial setup option is set to .
1. Resolution using the metric system (Unit: µm) Table 4.5.2.1A
emanledoM
01234567
6059-MSL
50.01.02.05.012 01001
2. Resolution using the inch system (Unit: inch) Table 4.5.2.1B
emanledoM
01234567
6059-MSL
200000.500000.10000.20000.50000.1000.5000.500.
.tinUgnirusaeMhcaefognittestluafedehtwohsserugifdedahsehT:1eto
N
fosnmulocehtnisnoituloseR:2etoN “ 0” .snacs23morfdeniatboebnachcihwesohtwohs
htiwsnmulocehtnisnoituloseR “ 1” .snac
s61morfdeniatboebnachcihwesohtwohs
tnemerusaemafotigidtnacifingistsaeleht,8ot1neewtebteserasnacsforebm
unehtfI:3etoN
.2ro,1,0.oNottessinoitulosererehwtuodeknalbyllacitamotuaeblliw
.ycaruccagnirusaemehtecudernetfoyamnoituloseregralootagnittestahtetoN:4etoN
forebmunehttes,eesottluciffiddnatcatniylesolcerastig
iddeyalpsidehterehW
.0b:edomputescisabehtninoitcnuftigidhtdnasuohtehtkramrostigidtuo-knalb
IMPORTANT Changing the resolution will cancel all the clibration values (HIGH CAL and LOW
CAL), offset value, mastering, abnormal value eliminating limits, GO/NG judgment
criteria, reference value, and setup values for the automatic workpiece detection.
Therefore, changing of the resolution should be carried out first.
b. Setting the number of blank-out digits (Guidance: )
Here, set the number of blank-out digits for measurements to be displayed in the display
unit. This blank out does not apply to the output to RS-232C interface, printer, Digimatic
output unit, and the display of setup value.
: No blank out (all digits are displayed) →
: The least significant digit is blanked out. →
: The least significant two digits are blanked out. →
(Default setting is .)
PROG
Step 1: Each time the key is pressed the displayed
figure changes in ascending order: → →
→ .
While the figure to be set is flashing, press the
ENT
key.
After accepting the specified value, the display proceeds to the setup stage of the
next item.