4.5.3.8 F7: Automatic workpiece detection setting
Set the conditions for automatic workpiece detection here.
Select between 0 (no automatic workpiece detection) and 999 measurements, and select
between 0 to 9999 ms for the invalidation period.
In this example assume the following:
Number of measuring times =1, Invalidation period = 100 ms (0.1 sec), Lower detection limit
= 12.2 mm, Upper detection limit = 12.8 mm.
PROG
Step 1: The previously set data output condition is
flashing.
1
PROG
Enter “1” as the number of measurements.
PROG
Step 2: Press the
ENT
key.
The operation automatically proceeds to the
invalidation period setting.
Step 3: Set the invalidation period to 100 ms.
00
1
PROG
PROG
Step 4: Press the
ENT
key.
The operation automatically proceeds to the lower
detection limit setting.
Step 5: Set the lower detection limit to 12.2 mm.
.2
12
PROG
PROG
Step 6: Press the
ENT
key.
The operation automatically proceeds to the upper
detection limit setting.
Step 7: Set the upper detection limit to 12.8 mm.
.8
12
PROG
PROG
Step 8: Press the
ENT
key.
The operation automatically proceeds to the group
judgment setting.