3 - 18
No. 99MBC032A
3.2.11 Reference value
In the basic setup the following conditions can be set.
a) Whether the target value of GO/NG judgment is being copied to the reference value. If
this is selected, the setup guidance for the reference value will not be displayed.
b) It is also possible to set so that tolerance judgment can take place in the ready state.
If the reference value is being set the deviation value (Measured data-reference value) will
be output for the RS-232C interface and the printer if single-run measurement or continu-
ous-run measurement is performed.
3.2.12 Data output conditions
In single-run measurement or continuous-run measurement, measured data can be output-
ted for each measurement if ±NG occurs, or at given intervals to the RS-232C interface,
printer, or Mitutoyo Digimatic Output Unit interface.
C.RUN
RUN
C.RUN
RUN
C.RUN
RUN
Data output condition Printer
RS-232C
DCU
Remark
The periodical output timer can be set
0
1
2
7
——
8
9
6
The periodical output timer can be set5
4
The periodical output timer can be set3
: Outputted for each measurement if or key, etc., is pressed.
: Press the or key to trigger the measurement. The measured data will be outputted if it falls on GO.
: Press the or key to trigger the measurement. The measured data will be outputted if it falls on ±NG.
: No output will be made.