A SERVICE OF

logo

6 - 22
No. 99MBC032A
6.2 Digimatic Output Unit interface
With the Digimatic Output Unit interface the LSM can be connected to the Digimatic Data
Processor (DP-1HS, etc.) which uses the Mitutoyo-original data format for easy data collec-
tion and processing.
6.2.1 Method of use
1) In the B6 mode of the basic setup, make the settings for the Digimatic Output Unit interface.
: The Digimatic Output Unit interface is valid.
Set the data output conditions in the function setup.
Data output condition Printer
RS-232C
DCU
Remark
The periodical output timer can be set
0
1
2
7
——
8
9
6
The periodical output timer can be set5
4
The periodical output timer can be set3
: Outputs data for each measurement.
: Performs measurement and outputs data when a GO measurement results.
: Performs measurement and outputs data when a ±NG measurement results.
: No output
2) Data output
a. When a data send request comes from the external equipment such as the Digimatic Data
Processor to the LSM, the measurement data will be outputted through this interface by
the following timing.
When the
DATA
key on the Digimatic Data Processor is pressed
When the foot switch being connected to the Digimatic Data Processor is pressed.
When an REQ signal is inputted from other external equipment.
b. When the measurement is initiated by pressing the
RUN
key or “R” command from the
RS-232C interface, the measurement data will be outputted according to the data output
conditions being set.