6. INTERFACE UNIT
TIP 1. To ensure compatibility with the LSM-9506, this system ignores the following
commands without treating them as ER6, but uses “OK”.
Memory switch command (UP0 uvxyz, UP1 uvxyz, UP2 uvxyz)
The MNL command is assumed to be identical to the MNH command.
2. If the setup data is “0”, it allows the corresponding function to be disabled. To set
“0” as a numerical value, use “0.0”.
Example 1 “SET, OF0” ................... Offset function is disabled.
Example 2 “SET, OF0.0” ................ Offset function is enabled.
3. Setup data or measured data is denoted by ±ddd.dddd.
The first (most significant) digit is a sign. For commands that do not specifically
designate the polarity, such as the calibration command, only a positive sign is
accepted. However, these positive signs do not have to be specified.
Any decimal place is selectable.
An integer can be entered without using a decimal point.
4. For making the setup operation of the measurement interval easier, use the
measurement interval number as shown below. For the correspondence between
the actual number of scans and the measurement interval numbers see the table
below.
Measurement interval number Number of scans for averaging
16 times
64 times
256 times
1
2
3
Measurement interval
0.01 sec
0.04 sec
0.16 sec
512 times4 0.32 sec
5. “Pp” command can be appended to the following commands to be treated as a
single command: “SET”, “R”, “CR”, “STAT”, “RP”.
Example: P0, R
P6, STAT