6. INTERFACE UNIT
,PR r
,PRT ttt
,SMP nnn
,(MAX,MIN,RNG,AVG)
,GN nn
,GLL±ddd.dddd
,GLH±ddd.dddd
,(GMX,GMN,GAG,GRG)
Data output conditions
Periodic data output conditions
Periodic output timer
Number of sample measurements
Setting the statistical item for sample measurement
(Maximum value, minimum value, range, mean)
Group sizes subject to judgment *2
Lower tolerance limit for group judgment
Upper tolerance limit for group judgment
Statistical item for group judgment
(Maximum value, minimum value, mean, range)
OK
Item
Reception command
Response command
(DATA FORMAT)R
CR
CL
D
ST
NST
(DATA FORMAT) *4
OK
(DATA FORMAT)
OK
OK
Start of single-run measurement
Continuous-run
measurement
Request of measurement data
R
STOP
No response command
(DATA FORMAT)
Zero-run measurement
Measurement start
command
Measurement stop
command
Measurement start
command
Measurement stop
command
Statistical processing
calculation
Performs statistical
processing
Does not perform
statistical processing
Current program only
All programs
Erasing the statistical
processing memory
MC
MCAL
OK
Request of statistical processing results
STAT (STAT FORMAT)
Condition setting of automatic workpiece detection *2
• Number of measurement times
• Invalidation period
• Lower detection limit
• Upper detection limit
Request of conditions list for automatic workpiece
detection
RA (RA FORMAT)
*1: • Do not use the SHL command unnecessarily unless there is a need to change the threshold voltage because of the
*1: • workpiece property, since it adversely affects the measuring accuracy.
*1: • In order to restore the standard threshold voltage, enter the following command:
*1: • “SHL50” Resets to 50%, which is the standard.
*1: • “STR” Records the value in memory.
*1: • The SHL command will be effective if the transparent measurement is selected in the basic setup mode.
*2: Designating any command of the functions which are set to “Not used” in the basic setup will result in ER6.
*3: If “Copying the target value to the reference value” is specified in the basic setup, designating this command results
*1: in ER-6.
*4: Responds with measurements according to the data output conditions.
Automatic workpiece detection control *2
AUT, S OK
AUT
,N mmm
,D tttt
,L±ddd.dddd
,H±ddd.dddd
OK
*2