![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/7/5f/75fcea55-90dc-4eca-9bdd-19ea4e0363bc/75fcea55-90dc-4eca-9bdd-19ea4e0363bc-bgc.png)
4 FONIX FP35 Hearing Aid Analyzer
[NEXT] [BACK] Moves back and forth between screens that are in a sequenc-
es. Examples include menus and the real-ear measurement
screens.
[∧, ∨] Moves the cursor up and down in a menu screen and in the
Audiogram Entry screen. Adjusts the level of the source ampli-
tude in a measurement screen.
[< , >] Cycles through available menu choices, adjusts the frequency
selection in a single tone measurement, and adjusts the ampli-
tude levels in an audiogram in the Audiogram Entry screen.
[START/STOP] Starts or stops the current selected function.
[OPERATE] Powers up and down the analyzer. When powered down, the
LCD display is turned off and all data is cleared, but the main
power is not turned off. The green LED will flash once every 3
seconds.
[RESET] Interrupts any current running measurement and returns the
FP35 to the Opening Screen. [RESET] will not clear leveling or
calibration, but it will clear any data you have collected.
[HELP] Pops up a help window containing instructions for the current
screen.
[PRINT/FEED] Prints the current screen when you press and release this key.
Feeds the thermal paper when you press and hold this key.
1.2.3 Rear Panel Layout
The rear panel contains most of the external connections for the FP35 analyzer.
There are two versions of the FP35 rear panel. The new rear panel (Figure
1.2.3A) was added in February 2007. All analyzers manufactured before this
date (that haven’t had a hardware upgrade) will have the older rear panel pic-
tured in Figure 1.2.3B.
VIDEO
EARPHONESPEAKER CONTRAST RS232 PRINTER POWER
Figure 1.2.3A—New FP35 rear panel layout
VIDEO
EARPHONE SPEAKER
CONTRAST
RS232
PRINTER KEYBOARD POWER
Figure 1.2.3B—Old FP35 rear panel layout