Chapter 5 487
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
indicate this range is currently inactive.
Key Path:
Meas Setup
Saved State: All values for all ranges are saved in instrument state.
Factory Preset: For each range, the following settings are selected:
Video BW (Auto),
Sweep Time (Auto), Absolute Limit = −50 dBm, Peak Threshold = −90 dBm, Peak
Excursion
=6dB,andAbs Stop Limit Mode (Auto) is set to −50 dBm. The table
below defines the other parameter’s settings for the measurement. The
values for ranges 1 through 5 are based on the W-CDMA (3GPP) BTS
Spurious Emission measurement.
Remote Command:
There is no equivalent remote command.
Table 5-6 Range Table Default Settings
Range Start Freq (GHz) Stop Freq (GHz) Res BW (kHz)
1 1.920 1.980 1200
2 1.894 1.920 510
3 2.100 2.102 100
4 2.175 2.180 100
5 0.800 1.000 4000
6-20 1.5 2.5 Auto