3. BASIC OPERATIONS
3.2.7 Setting the machine origin using the AT700 series
With the AT700 series if the machine origin is once determined, that reference point
cannot only be maintained through the subsequent normal operations such as zero setting,
etc., but also be protected from being lost even when the power is off.
Use the following procedure to set the machine origin offset value to Parameter 33.
(1) Make sure that Parameter 33 is 0.000.
1. While pressing the key turn the main switch on the rear panel to ON.
2. Make sure that the display looks like the following.
3. Enter Parameter 33.
4. Make sure that the display shows 0.000, then turn the main switch on the rear panel
to OFF.
(2) Measure the machine origin offset value.
1. Turn on the counter power and display Coordinate 0.
No. 99MBE032A
3-11