units:
W (Watts)
DBM (dB)
DBUV (dBmV)
DBMV (dBmV)
Remarks:
DBM 0dB is equal to 1mW readout mode
W = Watts readout mode
V = Volts readout mode. This selection is automatically made when the
channel input configuration is set to External volts (EXT V).
DBUV = dBmV, 0dB is equal to 1mV in readout mode.
Query:
CHUNIT? <c>
Returned
String:
CHUNIT <c>,<units>
When the channel input configuration is set to External volts (EXT V), the
returned units are always volts, irrespective of what units have been set.
CONT Continue
GPIB SETUP
Syntax:
CONT
Remarks:
This command will allow the system to continue the startup sequence if
there are self test failures other than DSP errors.
Related
Commands:
STERR, START
CUR Cursor in Power vs. Time and Source Sweep modes
SYSTEM
Syntax:
CUR <cursor>,<fval>
cursor:
1 or 2
fval:
0.0 to 1440 minutes (24 hours) in Power vs. Time mode
In Source Sweep mode, Power sweep : –120.0 dB to 30.0 dB
In Source Sweep mode, Frequency sweep : 10.0 KHz to 122.0 GHz
Remarks:
In Power vs. Time mode, the fval parameter is in minutes. In Source Sweep
mode, the fval parameter is in dB or Hz for a power sweep or frequency
sweep respectivily.
Examples:
Power vs. Time:
Set cursor 1 to 30 seconds: CUR 1,0.5
Set cursor 2 to 12.5 hours: CUR 2,750
Source sweep:
Power Sweep, set cursor 1 to 11.5 dB: CUR 1,11.5
Frequency Sweep, set cursor 2 to 15.6 GHz: CUR 2,15.6GHz
ML2430A OM 6-29
GPIB OPERATION ML24XXA NATIVE COMMANDS