If the source used does not provide a blanking output, the blanking signal may be
disabled as follows: select System|Rear Panel|BNC. Select PORT until INPUT 1
(digital) is selected, then select TTL Level to alter the active state of the blanking
signal expected. Setting the TTL Level to LOW will allow the Source sweep to
progress without a Blanking signal. This is useful for third party sources or simple
VTO systems. If a Source Sweep is later selected which does provide a blanking
signal, remember to restore the polarity of this signal to HIGH or an incorrect dis-
play will result. Operating a source sweep which has BANDSWITCH blanking de-
lays in it without an appropriate blanking signal may lead to glitches in the
resulting measurement at the bandswitch points (simple VTO systems do not
usually have bandswitch points).
Source sweep operation mode includes the following display controls:
CHANNEL
[GRMD]
Select Channel 1 or Channel 2. The channel selected will be dis
-
played on the left of the screen, above the middle value of the y
axis.
DATA HOLD
[GRPIX]
This selects how the graph is displayed on the screen. Select from
NORMAL, Min&Max, Min, or Max.
With Min & Max selected, the minimum and maximum points for
each sample are shown. If Connect Points is ON (SYSTEM|Graph
-
ics|CONNECT), a vertical bar is drawn between these points. Min
displays only the minimum for this sample. Max displays only the
maximum for this sample.
MODE
[SRCMOD]
Source sweep mode: FREQUENCY or POWER.
START
[SRCSTFRQ
SRCSTPWR]
Sweep start frequency (MHz or GHz) or power (dBm)
4-24 ML2430A OM
SYSTEM MENU OPERATION
NOTE
If either Min&Max, Min,
or Max is selected, the
display will keep the
“old” data and may ap
-
pear stationary. The
DATA HOLD mode in
use is displayed on the
left of the screen, below
the middle value of the
y axis.
ML2430A Series Power Meter
BNC Cables
Anritsu 68
/
69000
S
ynthesized
S
weep
G
enerator
DIGITAL INPUT 1
SEQ SYNC OUT
ANALOG INPUT 2
HORIZ OUT
DWELL
IN
10 MHz
REF OUT
EXT
ALC
IN
HORIZ
OUT
V/GHz
OUT
SEQ
SYNC
OUT
10 MHz
REF IN
SERIAL
I/O
AUX I/O
NOTE: USE SLOW BLO FUSES ONLY
!
!
CAUTION
>18kg
HEAVY WEIGHT
CAUTION
DO NOT OPERATE
WITH POWER CORD
UNGROUNDED
220V
2.5A
110V
5A
SELECT
LINE
INPUT
90 - 132VAC
180 - 264VAC
48 - 440Hz
IEEE-488 GPIB
110
220
Figure 4-5. Source Sweep Mode Interconnection Example