A SERVICE OF

logo

DE display enable
Syntax:
DE
Remarks:
Return the display to normal operation after the display has been set in DD
mode.
Related
Commands:
DD, DF
DF Display disable
Syntax:
DF
Remarks:
Turns the display off to allow faster measurements to be taken.
Related
Commands:
DD, DE
DR Set GPIB address
Syntax:
DR<val><terminator>
val:
1 to 30
terminator:
EN
Remarks:
Changes the device address. The power meter default address is 13.
DY Duty Cycle
Syntax:
DY<val><terminator>
val:
duty cycle value in percent
terminator:
%, PCT, or EN
Remarks:
Sets the duty cycle to be applied to the input signal.
EMUL GPIB emulation mode
Syntax:
EMUL <mode>
mode:
ML24XX (Anritsu ML2430A Series native mode)
HP436A (Hewlett-Packard)
HP437B (Hewlett-Packard)
ML2430A OM 6-105
GPIB OPERATION HP 437B EMULATION COMMANDS