V = Sensor A zeroed status: '0' = Not zeroed, '1' = Zeroed.
W = Sensor B Zeroed status: '0' = Not zeroed, '1' = Zeroed.
X = GPIB trigger mode: '0' = TR0 hold ON, '1' = Free run.
Y = GPIB group trigger mode: '0' = GTO, '1' = GT1, '2' = GT2.
Z = Calibrator state: '0' = OFF, '1' = ON.
1 = GPIB DISP command status: '0' = OFF, '1' = ON.
2 = GPIB FAST status: '0' = OFF, '1' = ON.
STERR Returns results of POST or *TST?
DATA OUTPUT
Syntax:
STERR
Remarks:
Returns (<sp> = space): 'FLASH<sp>0xnnnn,CALDAT<sp>0xnnnn,PER-
SON<sp>0xnnnn,RAM<sp>0xnnnn,NON-
VOL<sp>0xnnnn,LCD<sp>0xnnnn,KBD<sp>0xnnnn,DSP<sp>0xnnnn/n'
FLASH checksum test: 0x0000 = Passed, 0xffff = Failed
CALDAT checksum test: 0x0000 = Passed, 0xffff = Failed
PERSONality data: 0x0000 = Passed, 0xffff = Failed
RAM read/write test: 0x0000 = Passed, 0xffff = Failed
NONVOL RAM test: 0x0000 = Passed, 0x0001 = Software version fail,
0x0002 = Current store fail, 0x0004 = Saved store fail, 0x0008 = secure
mode fail, 0xffff = read failure
LCD memory test: 0x0000 = Passed, 0xffff = Failed
KBD stuck key test: 0x0000 = Passed, 0xffff = Failed
DSP test: 0x0000 = Passed, else FATAL error
Related
Commands:
START, CONT
SYSLD Load saved setup store over the GPIB
DATA OUTPUT
Syntax:
SYSLD <store number>, <data length>, <binary data>
store
number:
1 to 10
data length:
number of bytes of binary data
binary data:
Saved data previously read from the meter using the SYSRD command
6-74 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION