MANUAL (manual push button trigger)
CONT (continuous)
Remarks:
This command is overridden by the TR0, TR1 and TR2 commands when
in READOUT or POWER vs. TIME mode. If TR3 is sent, the trigger source
reverts back to the previously selected type of triggering.
Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger
together on the same conditions without having to set up two sets of trig
-
ger data.
Query:
TRGSRC? <c>
Returned
String:
TRGSRC <c>,<source>
The TRG type commands return the trigger state of the selected channel if
the channel is ON. This depends on the settings of the “link triggers” flag
and the current mode of the sensor on the selected channel. If the channel
is OFF, the stored trigger state of the channel is returned.
TRGTYP Set Trigger type
TRIGGER
Syntax:
TRGTYP <c>,<type>
c:
1, 2 or 1&2
type:
RISE
FALL
Remarks:
Sets the control type of the trigger used when the source is set to either
INTA or INTB (internal A or B) in READOUT or POWER vs. TIME mode.
Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger
together on the same conditions without having to set up two sets of trig
-
ger data.
Query:
TRGTYP? <c>
Returned
String:
TRGTYP <c>,<type>
The TRG type commands return the trigger state of the selected channel if
the channel is ON. This depends on the settings of the “link triggers” flag
and the current mode of the sensor on the selected channel. If the channel
is OFF, the stored trigger state of the channel is returned.
TRGXTTL Set external trigger edge type
TRIGGER
Syntax:
TRGXTTL <c>,<type>
c:
1, 2 or 1&2
ML2430A OM 6-81
GPIB OPERATION ML24XXA NATIVE COMMANDS