SENSTL Set Sensor Settle Percentage SENSOR
Syntax:
SENSTL <s>,<val>
s:
A or B
val:
0.01 to 10%
Remarks:
Sets how long the system waits for the signal to settle. The value parame
-
ter is only used in DEFAULT measurement sensor mode. The settling time
allows some control over the tradeoff between speed, and the extent to
which a measurement has settled to its final value.
Query:
SENSTL? <s>
Returned
String:
SENSTL <s>,<mode>
SENTYP Return sensor information
SENSOR
Syntax:
SENTYP <s>
s:
A or B
Remarks:
This command returns information on the selected sensor in string format:
<sensor type>,<sensor serial>.
For example: Dual Diode, PBD16.
SRCMOD Source Sweep Mode
SYSTEM
Syntax:
SRCMOD <mode>
mode:
FREQ
POWER
Remarks:
Determines whether the voltage sweep applied to the V/Ghz analog input
on the rear panel is interpreted as a frequency or power sweep. The x axis
of the graph on the display will be labeled appropriately.
Query:
SRCMOD?
Returned
String:
SRCMOD <mode>
SRCSPFRQ Source Sweep Stop Frequency
SYSTEM
Syntax:
SRCSPFRQ <freq_value>[units]
freq value:
10 kHz to 122 GHz
6-70 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION