Commands:
TR1, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR1 Trigger immediate
GPIB TRIGGER
Syntax:
TR1 <c>
c:
1 or 2
Remarks:
Triggers a single reading which is added to the internal digital filter and the
updated filter power level is returned on the GPIB. The returned reading
differs depending on the operation mode:
Readout: 'O' command response
Pwr vs. Time: 'O' command response
Profile: 'OGBD' command response (binary graph data for example)
Source Sweep: 'OGBD' command response (binary graph data for example)
After a TR1 command the instrument returns to either TR0 (trigger hold)
or TR3 (trigger free run) mode depending on what it was previously set to.
Related
Commands:
TR0, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR2 Trigger with a settling delay
GPIB TRIGGER
Syntax:
TR2 <c>
c:
1 or 2
Remarks:
Triggers a fully ranged and settled reading which is returned on the GPIB
Bus. If averaging is set to ON, the average buffer will be cleared and filled
before the result is returned. The returned reading differs depending on the
operation mode:
Readout: 'O' command response
Pwr vs. Time: 'O' command response
Profile: 'OGBD' command response (binary graph data for example)
Source Sweep: 'OGBD' command response (binary graph data for example)
NOTE
TR2 in Profile and Source Sweep mode is not supported
and will revert to a TR1 type measurement.
If the channel is set to External Volts, TR2 is not supported
(as there is no averaging and settling) and will revert to a
TR1 type measurement.
ML2430A OM 6-77
GPIB OPERATION ML24XXA NATIVE COMMANDS