A-2 ML2430A OM
SYSTEM SPECIFICATIONS APPENDIX A
Instrument Ac
-
curacy:
< 0.5 %
Zero Set and
Drift:
< 0.5 % MA2420A; < 0.5 % MA2470A Series and
MA2440A Series. Percent of full scale in most sensitive
range, measured over one hour with maximum averag
-
ing after one hour warm up at constant temperature.
Noise:
< 0.5 % of full scale in most sensitive range, measured
over a one minute interval with maximum averaging,
two standard deviations at constant temperature after
one hour warm up, typical. MA2470A Series, 20 pW
typical.
1.00 mW Power Reference
Frequency:
50 MHz nominal
Output Level:
1.00 mW, ± 1.2%/year, ± 0.9% RSS,
traceable to National Standards
VSWR:
< 1.04
Connector:
Type N female
SENSOR / CHANNEL CONTROL
Operating
modes:
Readout: dual channel. RF power or voltage.
Power vs. Time: single channel graphic of readout data
over adjustable time interval. RF power or voltage.
Profile: single channel RF peak power graphic display
for analysis of repetitive pulse or transient waveforms.
Source Sweep: Single channel graphic display syn
-
chronized to an RF source.
Range Hold:
Current range or selectable 1 through 5. (6 with
Universal Power Sensor).
Averaging
Auto-averaging: Moving average increases averaging
at low power ranges.
Averaging Types: Auto, Manual (Moving, Repeat)
Manual Average Range: 1, 2, 4, ..., 512
Low-Level Averaging: Low, Medium, and High settings
apply post-average low pass filter to improve visibility
at high display resolution settings.
Limit Lines:
Fixed value high and low limits with audible, rear panel
TTL output, and/or visible Pass/Fail alarm indication.
Failure indication can be set to latch until cleared so
that a transient failure can be easily noticed.