A SERVICE OF

logo

state:
ON or OFF
Remarks:
This will link the trigger set-up between Profile mode and Readout mode
so that the sample delay and the gate width will agree. A change to the
trigger set-up in either Readout or Profile system set-up will affect either
display mode.
Query:
LINK?
Returned
String:
LINK <state>
LLIM Set Low limits
CHANNEL
Syntax:
LLIM <c>,<val>
c:
1 or 2
val:
Remarks:
Sets the low limit. The LLIMS command turns the limits on and off. It is not
necessary to enter the units as the limit value is checked against the dis-
played value.
Therefore, if the limits have been set for –10 dBm (LLIM 1, –10) and the
display units are subsequently changed from dBm to Watts, the system still
checks for the reading to rise above –10, even though the display units
type has been changed.
Related
Commands:
LLIMS
Query:
LLIM? <c>
Returned
String:
LLIM <c>,<val>
LLIMS Turn on/off low limits
CHANNEL
Syntax:
LLIMS <c>,<state>
c:
1 or 2
state:
ON or OFF
6-52 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION
Units Min Max
dBm –99.99 +99.99
dBmV –53.00 147.00
dBµV
7.00 207.00
Watts 0.0 50.0
Volts 0.0 20.0