5 - 16
No. 99MBC071A
5.3.6 Applied measurement on a stepped round bar
øA øB
In this example 10 stepped round bars are measured
and the results are statistically processed. If ±NG
measurement is obtained, it will be automatically
printed out.
In the figure at the right suppose the following:
øA: ø6
±0.01
mm
øB: ø10h7
0
-0.015
mm
• Basic setup
1. Set the resolution to 0.1 µm.
2. Specify the RS-232C port as the printer port.
• Function setup
metiputeS)dnuorgeroF(0margorP)dnuorgkcab(1margorP
tnemgeS2 2
rofsnacsforebmuN
gnigareva
215215
eulavtimilrewoL99.5589.9
eulavtimilreppU10.60.01
noitidnoctuptuoataD2 2
noitidnocrehtO)lecnac(0)lecnac(0
• Preparation for measurement
Press the
SHIFT
and
A.CL
/
M.CL
keys to clear all of the statistical memory, then press the
STAT
/
S.E
key to start statistical processing. If the statistical processing mode is entered,
the S.E. guidance indicator ( ) turns on.
PROG
BUSY
RUN
+NG
GO
-NG
• Measurement
Perform a single-run measurement for the A dimension
by program No.0 after setting the workpiece in place.
The measured data will be displayed and the GO/NG
judgment result will be outputted.
PROG
BUSY
RUN
+NG
GO
-NG
Change to Program No.1 for the B dimension to be
measured through single-run measurement.
The measured data will be displayed and the GO/NG
judgment result will be outputted.
PROG
BUSY
RUN
+NG
GO
-NG
Change the workpiece and repeat the same measure-
ments.
PROG
BUSY
RUN
+NG
GO
-NG
If the result is ±NG, it will be automatically printed out.