6 - 27
No. 99MBC071A
6. INTERFACE UNIT
6.2 Optional Interface
Into the option slot either of the Digimatic Code Output Unit, Second I/O Analog, BCD, and
GP-IB Interface Units can be mounted.
6.2.1 Digimatic Output Unit interface
With the optional Digimatic Output Unit interface the LSM can be connected to the
Digimatic Data Processor (DP-1VR etc.) which uses the Mitutoyo-original data format for
easy data collection and processing.
In addition, as this LSM has two interface units, two pairs of measurements can be collected
at a time, if the simultaneous measurement is specified.
6.2.1.1 Method of use
1) Installation and setup of the interface
Install the interface unit in the Display Unit. For the method of installation, refer to
Section 6.3 “Installing the Optional Interface Unit”.
In the B6 mode of the basic setup, make the settings for the Digimatic Output Unit
interface.
:Uses only the OUTPUT-1 from the two interface units.
Select this mode for single measurement.
:Uses the OUTPUT-1 and OUTPUT-2 interface units.
Select this mode when two Digimatic Output Units are used for simultaneous
measurement.
Set the data output conditions in the function setup.
Data output condition Printer
RS-232C
GP-IB
DCU
Remark
The periodical output timer can be set
0
1
2
7
——
8
9
6
The periodical output timer can be set5
4
The periodical output timer can be set3
: Outputs data for each measurement.
: Performs measurement and outputs data when a GO measurement results.
: Performs measurement and outputs data when a ±NG measurement results.
: No output