A SERVICE OF

logo

4 - 29
No. 99MBC071A
4. SETTING UP THE MEASURING CONDITIONS
2) Moving average (Guidance: )
PROG
Different in the setup guidance ( ) and the
number of scans for averaging (between 32 and
2048), however, the setup method is same with
the arithmetical average.
NOTE 1. A larger number of scans for averaging will improve the repeatability.
If measuring time permits, set the greatest number of scans for averaging pos-
sible.
2. If the number of scans for averaging is set to between 1 and 4, the scan signals
will be thinned for the measurement. This results in a measurement interval of
0.002 to 0.003 second.