6 - 20
No. 99MBC071A
(14) SET
(a) Format: SET
Segment specification *1 , SG sssssss
Edge specification *1 , SG aaa-bbb
Measurement interval number at arithmetical average , M m
Number of scans for arithmetical average , MN nnnn
Measurement interval number at moving average , MS m
Number of scans for moving average , MNH nnnn
Lower abnormal limit *2 , EL±ddd.dddd
Upper abnormal limit *2 , EH±ddd.dddd
Abnormal count value *2 , CNT aaa
Lower limit *1 , LL±ddd.dddd
Upper limit *1 , LH±ddd.dddd
, L1±ddd.dddd
Multi-limit selection value *1 , L2 • • • •
, L6±ddd.dddd
Target value *1 , N±ddd.dddd
Lower tolerance limit *1 , LO±ddd.dddd
Upper tolerance limit *1 , UP±ddd.dddd
Reference value *3 , REF±ddd.dddd
Scale value , SCLc
Positive offset *4 , OF±ddd.dddd
Negative offset *4 , OM±ddd.dddd
Positive zero-set *4 , ZERO+
Negative zero-set *4 , ZERO–
Data output condition *5 , PRr
Periodic data output timer , PRT ttt
Number of sample measurements , SMP nnn
Sample measurement • Maximum value *6 , MAX
• Minimum value *6 , MIN
• Range *6 , RNG
• Mean *6 , AVG
Group size subject to judgment *2 , GN nn
Lower tolerance limit of group judgment *2 , GLL±ddd.dddd
Upper tolerance limit of group judgment *2 , GLH±ddd.dddd
Statistical items for group judgment: Maximum value *2, 6 , GMX
: Minimum value *2, 6 , GMN
: Mean *2, 6 , GAG
: Range *2, 6 , GRG
(b) Description: This sets the measuring conditions.
• Each of the commands that follow the SET command must be delimited by a
comma (,).
• A command which doesn’t need a setting change can be eliminated.
• Approximately 0.5 second is required for this command to be processed.