6 - 18
No. 99MBC071A
(5) SG sssssss, SG aaa-bbb
(a) Format: SG sssssss (sssssss: SEG No. Number of digits should be between 1 and 7.
Duplicated number must not be specified.)
SG aaa-bbb (aaa: start edge, bbb: finish edge. The range is between 1 and 255
for both edges. However, aaa should not be identical to bbb.)
(b) Description: Setting the segment (measuring position).
Two types of setting are available; segment specification and edge specification.
(c) Example: Reception command SG2 Response command OK
Reception command SG2-65 Response command OK
(d) Supplement: • Segments and edges should be set in the basic setup.
• sssssss can be set with 7 digits or less.
Ex.) SG 1234567, SG 24, SG3, etc.
• aaa and bbb should be set within 3 digits.
Ex.) SG1-2, SG 2-33, SG 111-255, etc.
The order of the start edge and finish edge can be reversed.
(6) M
(a) Format: M m (m: measurement interval number, 1 to 4)
(b) Description: Set the averaging method to the arithmetical average, and set a number between
1 and 4 for the representative measurement interval numbers. With these
measurement interval numbers compatibility with the conventional models is
ensured.
(c) Example: Reception command M4
Response command OK
(d) Supplement: Relationship between the measurement interval number, measurement interval,
and number of scans for averaging is as follows:
Measurement interval number Number of scans for averaging
16 times
64 times
256 times
1
2
3
Measurement interval
0.02 sec
0.08 sec
0.32 sec
512 times4 0.64 sec
(7) MN
(a) Format: MN nnnn (nnnn: Number of scans, between 1 and 2048)
(b) Description: Set the averaging method to the arithmetical average, and specify the number
of scans to nnnn (2
n
, where n= 0 to 11).
(c) Example: Reception command MN 1024
Response command OK
(8) MS
(a) Format: MS m (m: Measurement interval number, between 2 and 4)
(b) Description: Set the averaging method to the moving average, and specify a measurement
interval number to ensure compatibility with conventional models.
(c) Example: Reception command MS4
Response command OK
(d) Supplement: Relationship between the measurement interval number, measurement interval,
and number of scans for averaging is as follows: