4 - 28
No. 99MBC071A
4.5.3.2 F1: Setting the measurement interval (measurement time)
Use this function to set the measurement interval. This measurement interval should be set
according to the arithmetical average and moving average, whichever is specified in the basic
setup.
1) Arithmetical average (Guidance: )
PROG
Step 1: The previously set number of scans for averaging
is displayed. Select between 1 and 2048 times .
The relationship between the number of scans for
averaging and measurement intervals are shown in
the table below.
Relationship between the number of scans for averaging and measurement intervals (measure-
ment times)
Number of scans for
averaging
Measurement intervals (measurement time)
Arithmetical average Moving average
1st measurement
2nd and subsequent
measurements
1
—
2
—
4
—
8
16
32
64
128
256
512
1024
2048
—
—
0.02 sec
0.02 sec
0.02 sec
0.02 sec
0.02 sec
0.02 sec
0.02 sec
0.04 sec
0.08 sec
0.16 sec
0.32 sec
0.64 sec
1.28 sec
2.56 sec
0.04 sec
0.02 sec
0.01 sec
0.005 sec
0.0025 sec
0.0013 sec
0.08 sec
0.16 sec
0.32 sec
0.64 sec
1.28 sec
2.56 sec
Each time the key is pressed, the setup option changes in the following order:
→ → → → → → → → → →
→ → . For this example select 512 times.
PROG
Select 512 times.
PROG
ENT
Step 2: Press the
ENT
key to save the setup data in
memory.
The operation automatically proceeds to F2:
Setting the GO/NG judgment criteria.