6. INTERFACE UNIT
(c) Example: Reception command SET, SG2, M4, LL 5.988, LH 6.010, REF6.000,
SCL1, OM 0, PR 3, PRT10
*1: Select either setup method in the basic setup.
*2: This is valid only if the function is specified in the basic setup.
*3: This is valid only if the “Copying the target value to the reference value” is specified in
the basic setup.
*4: Set the reference gage on the Measuring Unit before sending this command. These
commands will spend approximately 1 second for processing. The settings of each
function are as follows:
Positive offset: Offsetting in the positive (0) direction.
Negative offset: Offsetting in the negative (1) direction.
Positive zero-set: Zero-setting in the positive (0) direction.
Negative zero-set: Zero-setting in the positive (1) direction.
*5: “PRr” is used to set the data output conditions for the RS-232C (printer)/GP-IB or
Digimatic Output Unit. If the PR number is 1, 3 or 5, it is possible to set the periodic
output timer, and the data output interval can be selected from 0 (for each measure-
ment) and between 1 and 999 seconds.
The PR numbers and the data output conditions have the following relationships.
Data output condition
(PR No.)
Printer
RS-232C
GP-IB
DCU
Remark
The periodical output timer can be set
0
1
2
7
——
8
9
6
The periodical output timer can be set5
4
The periodical output timer can be set3
: Outputs data for each measurement.
: Performs measurement and outputs data when a GO measurement results.
: Performs measurement and outputs data when a ±NG measurement results.
: No output
*6: Only one of these statistical items can be specified.
(15) R
(a) Format: R
(b) Description: If the number of samples is set between 1 and 999, this command executes
single-run measurement and transmits the measurement result in conformity
with DATA FORMAT as the response command.
(c) Example: Reception command R
Response command P0, 12.3456