A SERVICE OF

logo

6 - 39
No. 99MBC071A
6. INTERFACE UNIT
Continuous-run measurement with term specification
(SHIFT+)
PROG. 0 ~ 4
GO/NG result output
T
SO
T
IN
RESET
RUN
ACK
STB
T
AD
T
RA
T
AD
T
DO
T
W
T
BO
T
BC
T
BC
Timing of HOLD input (at continuous-run measurement)
(SHIFT+)
PROG. 0 ~ 4
GO/NG result output
T
SO
T
IN
RESET
RUN
ACK
STB
HOLD
T
IN
T
W
T
AD
T
RA
T
AD
T
DO
T
DO
or more
T
IN
T
BO
T
BC
T
DO
T
IN
Response time
Signal Response time Description
Input time
Acceptance time
Strobe length
Data setup time
Termination time
Single-run measurement
Zero-run measurement
Duration of continuous-run measurement
T
IN
TAD
TSO
TDO
TRA
TBO
TBC
10 ms or more
10 ms or less
Depends on the number of scans
(Refer to the following table.)
Input process timeT
W 10 ms or more
115 µs
20 ms or less
(Measurement interval) + 10 ms or less
(Measurement interval) x n + 10 ms or less
n = Number of measuring times
(Measurement interval) ± 5 ms or less
Strobe length: TSO
Number of scans for averaging Strobe length
0.3 ± 0.1 ms
2 ± 0.2 ms
20 ± 0.2 ms
1 ~ 8 times
16 ~ 128 times
256 ~ 2048 times