6. INTERFACE UNIT
6.2.2 Second Analog I/O Interface
This interface deals with two pairs of GO/NG judgment result output, one set of analog
output and control input. This is suitable for simultaneous measurement, multi-limit selection,
and group judgment operations.
6.2.2.1 Method of use
1) Installation and setup of the interface
• Install the interface unit in the Display Unit. For the method of installation, refer to
Section 6.3 “Installing the Optional Interface Unit”.
• In the basic setup make the settings according to the purpose of measurement. For detail
refer to Section 4.1 “Basic Setup”.
• In the function setup make the settings according to the purpose of measurement. For
detail refer to Section 4.5 “Setting Up the Functions”.
NOTE This interface has jumper pins (for short-circuiting between Pins 1 and 2 of JP1) on
the upper surface of the board. However, never modify the existing setting, since this
is for future expansion.
6.2.2.2 Name of each part
18
36 19
1
I/O ANALOG I/F
I/O0VALG FG
TP1
TP2
TP3
31
JP1
Connector
Analog output connector
I/O Interface connector
Connector 1 pin mark ( )
Jumper switch: JP1
fognitsisnoCytitnauQ
ecafretniO/IgolanAdnoceS1
rotcennoC1
elbacgnitcennoC1