6.1.1.2 Terminal names
Terminal No. Signal name Function I/O direction
Frame ground (connected to the casing)
Used for connecting the shielded wire of I/O signal cables
Output of measurement condition (status)
Turned out “H” level in the event of “Err-0”
• GO/NG judgment result output (GO)
• With the basic setup, this can be changed to strobe signal (STB) or
measurement in-progress signal (ACK) output.
GO/NG judgment result output (+NG)
GO/NG judgment result output (-NG)
Frame ground (connected to the casing)
Used for connecting the shielded wire of I/O signal cables
Analog voltage output
Offset input
Can be changed to hold (HOLD) by the basic setup.
Input of trigger command of single-run measurement
Can be changed to a trigger for continuous-run measurement or
continuous-run measurement with term specification by the basic setup.
• Input of CLEAR command, same as the key
A1
A2
A3
A4
A5
B1
B2
B4
B5
B6
FG
STS
GO
+NG
–NG
FG
ALG
OFFS
RUN
RES
Out
Out
Out
Out
Out
Out
In
In
In
B3 0V
GND Digital ground
Common ground terminal of both output (A2 thru A5) and input (B4 thru B6)
A6 GND
0V output of analog voltage output Out
C
6.1.1.3 Input/output equivalent circuit
(1) Input circuit
2.7K
Input signal
(OFFS, RUN, RES)
Connected to the inside (+24V)
Input low-level signals between 0 and 1 V. Generally drive this circuit with an open
collector-type transistor.
Maximum current drawn from the input signal terminal is 12 mA.