4 - 38
No. 99MBC071A
IMPORTANT How to use the offset function
1. To obtain an offset value, it is necessary to set up the reference gage in place
(the offset value is a compensation value determined from the measurement of
the reference gage). This offset setup takes about 1 second.
PROG
2. If the existing setup value is applied, it is not
necessary to carry out the offset. To force the
offset operation using the same data, move the
highlighted digit place with the key. This
makes the offset carried out, since the system
judges the data is changed.
For this operation press the
MASTER
/
OFFSET
key in the
ready state. With this single key operation offset
will be carried out.
3. So that the maximum displayable range is not exceeded during measurement, the
offset value must be set well within the maximum value shown in the table below.
If measured data exceed the maximum value, “9999999” will be displayed.
(noituloseR µ )m)mm(eulavmumixaM)E(noituloseR)E(eulavmumixaM
50.0/20.0/10.0 ± 99999.98
/200000./100000.
500000.
± 999999.8
5.0/2.0/1.0 ± 9999.998
/50000./20000./10000.
500./5000./2000./1000.
± 99999.98
001/01/5/2/1 ± 999.9998
4. Precautions prior to modifying the unit system
Note that if an offset value exceeds the above described maximum value when
the unit system is changed from E to metric, the offset function will automatically
be reset.
(Example: If the integer part of the maximum value is restricted to 2 digits,
converting from 4E to 101.6 mm will exceed the limit.)