A SERVICE OF

logo

3 - 25
No. 99MBC071A
3. DISPLAYS AND KEY OPERATIONS
3.4.1 Description of key functions
90
+/-
C
SHIFT
STAT
S.E
SHIFT
SHIFT
READ
SHIFT
RUN
S.PR
PRINT
S.PR
PRINT
C.RUN
RUN
C
Key name In the ready state
In the display-latched state
At single-run measurement
At continuous-run measurement
At setup
Combined use with power-on
operation
Operation error
Operation error
Shift key
To enter the function indicated in
the upper portion of a double-
function key, such as the
key, hold down the
before pressing the key.
A foreground program number will
flash for about 10 seconds.
Cancels the error that occurred
when the power was turned on.
Cancels the latched state and
returns to the ready state.
Changes the program number
Operation error
Operation error
Operation error
Entry of +
(to set the light amount detecting
function) is valid when the function
setup item number flashes in the
function setup mode.
Performs single-run measurement
(even in the display-latched state).
Results in a single-run
measurement error.
Quits the measuring operation for
continuous-run measurement.
Operation error
Starts continuous-run measurement
(even in the display-latched state).
Quits the measuring operation for
continuous-run measurement
(same as ).
Operation error
Prints out the previous
measurement data.
Prints out the data currently
displayed in the display-latched
state.
Results in a single-run
measurement error.
Prints out the previous
measurement data in
continuous-run measurement.
Operation error
If the printer is active, prints out all
the statistical processing data and
clears the statistical memory.
If the printer is not active, results in
an operation error.
Operation error Operation error
Aborts the measurement and
returns to the ready state.
Operation error
Enters a decimal point.
Inverts the sign of the setup value.
Cancels the setup value or resets
it to initial value.
Cancels the error state.
+ power-on will enter the
initialization mode of the Display
Unit.
Enters the setup data.