A SERVICE OF

logo

3 - 17
No. 99MBC071A
3. DISPLAYS AND KEY OPERATIONS
3.2.12 Data output conditions
In single-run measurement or continuous-run measurement, measured data can be output-
ted for each measurement if ±NG occurs, or at given intervals to the RS-232C/GP-IB
interface, printer, or Mitutoyo Digimatic Output Unit.
C.RUN
RUN
C.RUN
RUN
C.RUN
RUN
Data output condition Printer
RS-232C
GP-IB
DCU
Remark
The periodical output timer can be set
0
1
2
7
——
8
9
6
The periodical output timer can be set5
4
The periodical output timer can be set3
: Outputted for each measurement if or key, etc., is pressed.
: Press the or key to trigger the measurement. The measured data will be outputted if it falls on GO.
: Press the or key to trigger the measurement. The measured data will be outputted if it falls on ±NG.
: No output will be made.
3.2.13 Automatic workpiece detection <Diameter detection method, Position detection
method>
Automatic workpiece detection is performed for continuous-run measurement, where
measurement starts with no specified workpiece present (Err-0), then proceeds to auto-
matic detection of the workpiece, followed by measurement repeated number of times. No
specified workpiece present (Err-5) also refers to the workpiece outside the upper and
lower detection limits.
Whether automatic workpiece detection is performed is specified in the basic setup mode.
If automatic workpiece detection is specified, the number of scanning times for detection
must be specified from among 1 and 16. Select 16 times if detecting precision workpieces.
If automatic workpiece detection is not specified, no further setting is necessary.
Automatic workpiece detection setup includes the number of measurement times, invalida-
tion period, upper and lower detection limits. Both the upper and lower detection limits
may not always need to be specified, however, they should be set for safety.
To exclude the measured data of such as chamfered portion of the workpiece, invalidation
period can be set within the range from 0.001 sec to 9.999 sec.
No workpiece
detected
Start of
measurement
Detection of
workpiece
Invalidation
elapses
Measured the specified
number to times