6 - 19
No. 99MBC071A
6. INTERFACE UNIT
Measurement interval number Number of scans for averaging
64 times
256 times
2
3
1st measurement interval
0.08 sec
0.32 sec
512 times4 0.64 sec
2nd measurement interval
0.02 sec
0.02 sec
0.02 sec
(9) MNH
(a) Format: MNH nnnn(nnnn: Number of scans, between 32 and 2048)
(b) Description: Set the averaging method to the moving average, and specify the number of
scans with nnnn. nnnn is 2
n
, where n= 5 to 11.
(c) Example: Reception command MNH 1024
Response command OK
(d) Supplement: MNL command is as same as MNH command
(10) STR
(a) Format: STR
(b) Description: Data that has been set by the RS-232C command will be erased from memory
if the power is off. To retain the data after the power off, use this command to
save the critical measuring conditions in memory. But the “ST” and “NST”
command will not be saved.
(c) Example: Reception command STR
Response command OK
(11) LOCK
(a) Format: LOCK
(b) Description: Locks the keyboard of this machine to prevent accidental operation.
To release this key lock state, use the UNLOCK command.
(c) Example: Reception command LOCK
Response command OK
(d) Supplement: Lock set by this command can not be released with key operation.
(12) UNLOCK
(a) Format: UNLOCK
(b) Description: Releases the key lock state and enables key operations again.
(c) Example: Reception command UNLOCK
Response command OK
(13) RP (RP FORMAT)
(a) Format: RP
(b) Description: This is used to confirm the setup contents, if the measuring conditions and
operating conditions set are received as the response.
(c) Example: Reception command RP command
Response command PROGRAM, SG 2, M3, LL D 5.988, LH 6.010,
REF6.000, SCL1, OF 0, PR 3, PRT0, SMP20, MAX,
ST