A SERVICE OF

logo

6 - 35
No. 99MBC071A
6. INTERFACE UNIT
b. Pin assignment for multi-limit selection (L1-L6)
Pin No.
Signal name
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
+5V
COM (IN)
PROG. 0
PROG. 2
PROG. 4
SHIFT
RUN
A-L1
A-L3
A-L5
B-L1
B-L3
B-L5
A-L7
A-L2
ERR. 0
COM (OUT)
CNT
I/O direction
(Internal power)
(IN)
IN
IN
IN
IN
IN
OUT
OUT
OUT
OUT
OUT
OUT
OUT
OUT
OUT
(OUT)
OUT
Pin No.
Signal name
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
GND
COM (IN)
PROG. 1
PROG. 3
B-L7
PRINT
RESET
A-L2
A-L4
A-L6
B-L2
B-L4
B-L6
A-L1
ACK
STB
COM (OUT)
FG
I/O direction
(Internal power)
(IN)
IN
IN
OUT
IN
IN
OUT
OUT
OUT
OUT
OUT
OUT
OUT
OUT
OUT
(OUT)
Applicable connector
57-30360 (or the equivalent product by DDK or Anphnor, etc.)
This is the standard accessory of this interface.
NOTE 1. Between Pin No.8 and Pin No.32, and between Pin No.15 and Pin No.26 are
connected internally, therefore they output the identical signals. Only use either of
the pins.
2. In simultaneous measurement results from the program Nos.0 to 4 will be output-
ted to A-L1 through A-L7, and that from the program Nos.5 to 9 will be outputted
to B-L1 through B-L7, respectively.
3. In single measurement the identical results will be outputted to A-L1 through A-L7
and B-L1 through B-L7, respectively.
4. The group judgment function can not be used for multi-limit selection.