No. 99MBC071A
4.5.3.3 F2: Setting the GO/NG judgment criteria
Set the GO/NG judgment criteria according to the tolerance judgment method: (Lower limit
value + Upper limit value), (Multi-limit selection: 7 stages), and (Target value + tolerance),
whichever is specified in the basic setup. If Using the abnormal value elimination function
has been specified, the abnormal limit values should be set prior to setting the GO/NG
judgment criteria.
In this example assume that the machining target value is 12.5
+/- 0.01
mm, and that all the
abnormal limits (lower and upper) and GO/NG judgment criteria are canceled (set to 0).
1) Setting the abnormal limit values
Set as follows: Lower abnormal limit = 12.48 mm, Upper abnormal limit = 12.52 mm,
Abnormal value count = 3.
PROG
.4
12
8
PROG
Step 1: The previously set lower abnormal limit is
displayed.
Enter 12.48.
PROG
Step 2: Press the
ENT
key.
The setup data will be saved in memory and
operation automatically proceeds to the setting for
the upper abnormal limit.
Step 3: Enter 12.52, which is the setup data for the
upper abnormal limit.
.5
12
2
PROG
PROG
Step 4: Press the
ENT
key.
The setup data will be saved in memory and
operation automatically proceeds to the abnormal
value count setting.
3
PROG
Step 5: Enter 3, which is the setup data for the
abnormal value count.
PROG
Case of
(Lower limit value and Upper limit value)
Step 6: Press the
ENT
key.
The setup data will be saved in memory
and operation will automatically proceed to
the GO/NG judgment criteria setting.
The method of GO/NG judgment varies
with the contents of the basic setup.
PROGPROG
Case of (Multi-limit selection) Case of (Target value + tolerance)